Table T12. Magnetic survey of the shield oven.
Distance
(cm)
x (up)
(mG)
y (starboard)
(mG)
z (into magnetometer)
(mG)
Total
(mG)
Comment
0.0 138.00 –6.00 –4.00 138.19 Mouth of oven
5.0 110 10.00 –3 110.49 Heating chamber
10.0 48.00 4.00 –0.80 48.17 Heating chamber
15.0 17.90 1.10 –0.70 17.95 Heating chamber
20.0 6.40 0.70 –0.80 6.49 Heating chamber
25.0 2.20 0.20 –0.40 2.24 Heating chamber
30.0 1.30 0.13 –0.18 1.32 Heating chamber
35.0 1.29 0.14 –0.20 1.31 Heating chamber
40.0 1.82 0.33 –0.72 1.98 Heating chamber
45.0 3.05 0.77 –2.41 3.96 Heating chamber
50.0 4.68 2.22 –5.87 7.83 Heating chamber
55.0 4.51 2.68 –4.23 6.74 Heating chamber
60.0 7.80 1.24 –0.66 7.93 End of the heating chamber
65.0 3.15 0.02 –0.13 3.15 Cooling chamber
70.0 0.51 0.03 –0.09 0.52 Cooling chamber
75.0 0.13 0.05 –0.08 0.16 Cooling chamber
80.0 0.16 0.04 –0.06 0.18 Cooling chamber
85.0 0.18 0.02 –0.07 0.19 Cooling chamber
90.0 0.20 0.01 –0.07 0.21 Cooling chamber
95.0 0.23 0.00 –0.07 0.24 Cooling chamber
100.0 0.26 –0.01 0.07 0.27 Cooling chamber
105.0 0.47 –0.20 0.07 0.52 Cooling chamber
110.0 4.29 0.14 2.48 4.96 Cooling chamber
115.0 56.10 2.00 18.20 59.01 Cooling chamber (but beyond where samples holder resides during cooling)
120.0 198.00 7.00 37.00 201.55 Cooling chamber (but beyond where samples holder resides during cooling)
125.0 393.00 20.00 91.00 403.89 Cooling chamber (but beyond where samples holder resides during cooling)
130.0 764.00 50.00 231.00 799.72 Cooling chamber (but beyond where samples holder resides during cooling)
134.0 884.00 58.00 317.00 940.91 End of cooling chamber