Figure F14. Magnetic moments for x-, y-, and z-axis measured for an empty discrete sample tray on the sample boat. Moment was measured after the tray was cleaned with window cleaner and alternating-field demagnetized at 80 mT. Drift correction has been applied, as well as tray correction that removes magnetization related to the sample boat but not the discrete sample tray.
Previous | Close | Next | Top of page