Table T12. Replicate analyses of an interstitial water sample to establish typical precision of Dionex ICS-3000 Ion Chromatograph for a typical anion (chloride) and cation (sodium), Expedition 317.
Analysis no. |
Chloride (mM) |
Sodium (mM) |
---|---|---|
1 | 326.9 | 278.8 |
2 | 327.7 | 278.4 |
3 | 326.3 | 279.4 |
4 | 327.4 | 279.3 |
5 | 327.9 | 277.9 |
6 | 327.1 | 279.3 |
7 | 326.5 | 278.0 |
Mean: | 327.1 | 278.7 |
Standard deviation: | 0.611 | 0.649 |
Coefficient of variation: | 0.002 | 0.002 |
Previous | Close | Next | Top of page