Table T3. Precision and accuracy for major and trace element analyses by ICP-AES, Expedition 329.
|
Wavelength accepted
|
BCR-2 (n = 24)
|
BAS-206 (n = 8)
|
Accepted international
|
Shipboard average
|
Uncertainty
|
SD (%)
|
Accepted average*
|
Shipboard average
|
Uncertainty
|
SD (%)
|
Major element oxide (wt%):
|
SiO2
|
251.611
|
54.1
|
54.6
|
0.65
|
1.19
|
49.29
|
49.4
|
0.86
|
1.75
|
Al2O3
|
396.152
|
13.6
|
13.5
|
0.39
|
2.92
|
14.46
|
13.8
|
0.15
|
1.06
|
Fe2O3(T)
|
239.563
|
13.4
|
14.0
|
0.26
|
1.87
|
15.48
|
14.5
|
0.13
|
0.89
|
MnO2
|
257.61
|
0.2
|
0.2
|
0.01
|
5.15
|
0.26
|
0.2
|
0.02
|
6.54
|
MgO
|
285.213
|
3.4
|
3.9
|
0.18
|
4.51
|
6.94
|
6.8
|
0.10
|
1.54
|
CaO
|
317.933
|
7.0
|
7.4
|
0.30
|
4.05
|
9.63
|
9.8
|
0.29
|
2.99
|
Na2O
|
589.592
|
3.3
|
3.4
|
0.25
|
7.33
|
2.72
|
2.8
|
0.05
|
1.79
|
K2O
|
766.491
|
1.7
|
1.8
|
0.09
|
4.69
|
0.16
|
0.2
|
0.06
|
25.75
|
TiO2
|
337.28
|
2.3
|
2.2
|
0.10
|
4.51
|
2.12
|
1.9
|
0.07
|
3.40
|
P2O5
|
214.914
|
0.4
|
0.4
|
0.06
|
13.25
|
0.20
|
0.2
|
0.05
|
19.92
|
Trace element (ppm):
|
Ba
|
493.409
|
683
|
540
|
98.37
|
18.22
|
44.7
|
14.3
|
36.97
|
258.73
|
Co
|
228.615
|
42
|
51
|
2.64
|
5.22
|
53.8
|
59.8
|
3.47
|
5.80
|
Cu
|
324.754
|
19
|
NR
|
—
|
—
|
67.9
|
58.6
|
8.61
|
14.68
|
Ni
|
231.604
|
12
|
11
|
3.62
|
32.79
|
51.2
|
49.2
|
2.32
|
4.72
|
Sc
|
361.383
|
33
|
33
|
1.08
|
3.29
|
46.9
|
50.2
|
0.54
|
1.08
|
Sr
|
407.771
421.552 |
346
|
342
|
13.24
|
3.87
|
100.1
|
99.5
|
2.48
|
2.50
|
V
|
292.401
|
416
|
411
|
8.18
|
1.99
|
469.0
|
455.0
|
6.04
|
1.33
|
Zn
|
213.856
|
127
|
132
|
5.93
|
4.49
|
192.3
|
125.6
|
3.49
|
2.78
|
Zr
|
343.823
|
188
|
179
|
12.89
|
7.19
|
124.5
|
112.6
|
2.90
|
2.58
|
* = based on multiple repeat runs of BAS-206 standard against known internationally accepted standards carried out at the National Oceanography Centre, Southampton, United Kingdom. n = number of replicates, SD = standard deviation. NR = not reported, — = not analyzed.