Table T7. Typical sequence of analyses in ICP-AES run, Expedition 330.

Analysis no. Identifier/Type
1 DRIFT 1 (BHVO-2)
2 BLANK - 1
3 BHVO-2 STD
4 BIR-1 STD
5 BCR-2 STD
6 DRIFT 2 (BHVO-2)
7 JGB-1 STD
8 JA-3 STD
9 SCO-1 STD
10 JR-1 STD
11 DRIFT 3 (BHVO-2)
12 MRG-1 STD
13 NBS-1C STD
14 Check
15 Check
16 DRIFT 4 (BHVO-2)
17 Sample
18 Sample
19 Sample
20 Sample
21 DRIFT 5 (BHVO-2)
22 Sample
23 Sample
24 Sample
25 Sample
26 DRIFT 6 (BHVO-2)
27 Sample
28 Sample
29 Sample
30 Sample
31 DRIFT 7 (BHVO-2)
32 BLANK - 2
33 Check
34 Check
35 Sample
36 DRIFT 8 (BHVO-2)
37 Sample
38 Sample
39 Sample
40 Sample
41 DRIFT 9 (BHVO-2)
42 Check
43 Check
44 BLANK - 3
45 DRIFT 10 (BHVO-2)

STD = calibration standard, check = check standard, DRIFT = drift correction standard. As many as 22 samples were included in a batch.