Previous   |    Close   |    Next

Table T5. Analytical conditions for major element analysis of glass beads, Expedition 331.

Element
line
Filter Crystal Peak
angle
(°)
Count
time
(s)
BG 1
(°)
Count
time
(s)
BG 2
(°)
Count
time
(s)
Detector
Si-Kα Out PET
108.946
40
106.100
10
111.250
10
PC
Ti-Kα Out LiF1
86.155
20
85.240
10
87.120
10
SC
Al-Kα Out PET
144.607
40
147.150
10
140.400
10
PC
Fe-Kα Out LiF1
57.535
20
58.180
10
56.900
10
SC
Mn-Kα Out LiF1
62.997
20
62.200
10
63.820
10
SC
Ca-Kα Out PET
45.154
40
43.650
10
 
 
PC
Mg-Kα Out RX25
38.804
40
40.750
10
36.900
10
PC
Na-Kα Out RX25
47.125
40
49.000
10
45.250
10
PC
K-Kα A 140 PET
50.632
40
49.200
10
 
 
PC
P-Kα Out PET
89.280
40
91.350
10
87.200
10
PC

Analyses performed on the Supermini (Rigaku) X-ray fluorescence spectrometer. BG = background angle. PC = flow-proportioned counter, SC = scintillation counter.