Previous   |    Close   |    Next

Table T4. Standard reproducibility and accuracy and precision estimates for gas chromatography analyses of CO2, H2O and S, Expedition 345.

Standard Date CO2
(wt%)
H2O
(wt%)
S
(ppm)
JP-1 24 Dec 2012 0.28 2.50 ND
  26 Dec 2012 0.33 2.48 ND
  3 Jan 2013 0.35 3.02 ND
  8 Jan 2013 ND 2.13 ND
  12 Jan 2013 0.26 2.61 ND
  13 Jan 2013 0.29 2.09 ND
  17 Jan 2013 0.29 2.44 ND
  28 Jan 2013 0.27 2.53 ND
  3 Feb 2013 0.25 2.62 ND
  8 Feb 2013 0.28 2.66 ND
Average (wt%/ppm): 0.289 2.507 ND
Preferred value (wt%/ppm): 0.280 2.540  
Standard deviation (wt%): 0.030 0.26 ND
Precision (%): 11 11 ND
Accuracy (%): 3 –1 ND
MRG-1 24 Dec 2012 1.09 0.80 ND
  26 Dec 2012 1.07 0.90 589
  3 Jan 2013 1.19 1.16 ND
  8 Jan 2013 ND 0.96 666
  12 Jan 2013 1.04 0.89 521
  13 Jan 2013 1.17 0.92 ND
  17 Jan 2013 1.18 0.82 ND
  28 Jan 2013 1.15 0.99 504
  3 Feb 2013 1.10 0.96 ND
  8 Feb 2013 1.15 0.94 ND
Average (wt%/ppm): 1.126 0.93 570
Preferred value (wt%/ppm): 1.070 1.02 605
Standard deviation (wt%): 0.053 0.099 74
Precision (%): 5 11 13
Accuracy (%): 5 –8 –6

ND = not determined.

Previous   |    Close   |    Next   |    Top of page