Table T3. MSCL sensor specifications.

Sensor
Calibration pieces used
Acceptable
departure from full
calibration values
Standard
sampling interval
(cm)
Sampling interval
for M0061 and
M0062 (cm)
Spatial
resolution (cm)
Full calibration
Calibration check
Gamma density Stepped Al/H2O (6, 5, 4, 3, 2, and 0 cm Al) Distilled water ±100 cps 2 1 1
P-wave velocity Distilled water Distilled water ±1 μs 2 1 1
Noncontact resistivity Saline fluids (35, 17.5, 8.75, 3.5, 1.75, and 0.35 g/L) 17.5 g/L saline fluid ±10 mV 2 1 2–3
Magnetic susceptibility 80 mm loop Impregnated resin Impregnated resin ±2%–4% 2 1 2–3
Magnetic susceptibility 90 mm loop Impregnated resin Impregnated resin ±2%–4% 4 4 3–4