Table T6. Analytical conditions for major element analysis of glass beads on the Supermini (Rigaku) X-ray fluorescence spectrometer, Expedition 348.

Element-
line
Filter Crystal Peak
angle
(°)
Count
time
(s)
BG 1
angle
(°)
Count
time
(s)
BG 2
angle
(°)
Count
time
(s)
Detector
Na-Kα OUT RX25 47,129 40 49,450 10 PC
Mg-Kα OUT RX25 38,802 40 35,450 10 41,550 10 PC
Al-Kα OUT PET 144,647 40 140,150 10 147,750 10 PC
Si-Kα OUT PET 108,980 40 106,000 10 119,900 10 PC
P-Kα OUT PET 89,360 40 91,550 10 PC
K-Kα A 140 PET 50,648 40 48,600 10 PC
Ca-Kα OUT PET 45,176 40 43,200 10 PC
Ti-Kα OUT LiF1 86,138 20 87,380 10 SC
Mn-Kα OUT LiF1 62,959 20 64,380 10 SC
Fe-Kα OUT LiF1 57,527 20 58,820 10 SC

PC = proportional counter, SC = scintillation counter. — = no data.