Previous   |   Close   |   Next

Table T10. Analytical conditions for hard rock inductively coupled plasma–atomic emission spectroscopy runs.

Element Wavelength (nm) Integration time per calculation point (s) Voltage
(V)
Mode Increment between points (nm)* N
Al 396.15 1 570 Gauss 0.003 5
Ba 455.40 4 610 Max 0.003 1
Ca 393.37 1 390 Gauss 0.003 5
Cr 267.72 4 820 Max 0.003 1
Fe 259.94 1 570 Gauss 0.003 7
K 766.49 1 990 Gauss 0.003 5
Mg 285.21 4 640 Max 0.003 1
Mn 257.61 4 560 Max 0.003 1
Na 589.59 1 660 Gauss 0.003 5
Ni 231.60 4 930 Max 0.002 1
P 178.23 4 990 Max 0.003 1
Si 251.61 1 580 Gauss 0.003 5
Sr 407.77 1 700 Gauss 0.003 5
V 292.40 1 890 Gauss 0.003 7
Y 371.03 4 620 Max 0.003 1
Zr 343.82 1 640 Gauss 0.003 5
Ti 334.94 1 620 Gauss 0.003 5
Nb 316.34 1 550 Gauss 0.003 5
Sc 361.38 0.5 620 Gauss 0.002 5

Notes: Reported settings are stored in the software. Entrance and exit slits are the same for each element, 20/15 nm, respectively. * = the interval between each of the calculation points in Gaussian (Gauss) mode or the calculation window that constitutes the single point in maximum (max) mode. N = number of calculation points.