Equipment
|
Status
|
Quality assessment
|
2G Enterprises 760R superconducting rock magnetometer
|
Used and functional
|
Magnetic moments for a 150 cm blank (air) after tray and background corrections is <2 × 10–10 Am2. Magnetic field in sensor region is <8 nT.*
|
AGICO KappaBridge KLY-4S susceptibility meter
|
Used and functional
|
Used to measure bulk susceptibilities.
|
DTECH D-2000 alternating-field (AF) demagnetizer
|
Not used
|
Undetermined.
|
Schonstedt TSD-1 thermal demagnetizer
|
Used and functional
|
Thermal demagnetization experiments were conducted on 13 discrete samples.
|
ASC IM-10 impulse magnetizer (Serial 01971)
|
Tested and functional
|
Undetermined.
|
Applied Physics portable fluxgate magnetometer (model 520)
|
Tested and functional
|
Resolution <1 nT.*
|
Molspin minispin magnetometers (two)
|
Not used
|