Table T7. Handheld X-ray fluorescence measurements of BCR-2 standard by using rock standard calibration, Expedition 345.
|
Preferred value*
|
Expedition 335
|
Expedition 345
|
Calibration
|
Accuracy (%)
|
Calibration (N = 5)
|
Precision (%)
|
Accuracy (%)
|
Major element (wt%):
|
SiO2
|
54.1
|
48.14
|
–11.02
|
52.61
|
1.52
|
–2.75
|
TiO2
|
2.26
|
1.77
|
–21.62
|
2.27
|
0.92
|
0.38
|
Al2O3
|
13.5
|
8.9
|
–34.07
|
9.4
|
6.2
|
–30.37
|
FeOT
|
12.28
|
12.41
|
1.07
|
12.10
|
0.18
|
–1.48
|
MnO
|
0.2
|
0.22
|
8.86
|
0.19
|
1.21
|
–5.88
|
MgO
|
3.59
|
BDL
|
—
|
BDL
|
—
|
—
|
CaO
|
7.12
|
7.03
|
–1.33
|
6.77
|
0.87
|
–4.92
|
Na2O
|
3.16
|
—
|
—
|
—
|
—
|
—
|
K2O
|
1.79
|
1.12
|
–37.43
|
0.97
|
3.08
|
–45.81
|
P2O5
|
0.35
|
BDL
|
—
|
BDL
|
—
|
—
|
Total:
|
98.35
|
|
|
|
|
|
Trace element (ppm):
|
Ba
|
677
|
BDL
|
—
|
BDL
|
—
|
—
|
Sr
|
340
|
240
|
–29.41
|
353
|
2.98
|
3.82
|
Zr
|
184
|
151
|
–17.80
|
178
|
3.44
|
–3.26
|
V
|
416
|
150
|
–63.94
|
535
|
7.49
|
28.61
|
Cr
|
18
|
BDL
|
—
|
BDL
|
—
|
—
|
Co
|
37
|
BDL
|
—
|
BDL
|
—
|
—
|
Ni
|
18
|
BDL
|
—
|
BDL
|
—
|
—
|
Cu
|
21
|
BDL
|
—
|
BDL
|
—
|
—
|
Zn
|
127
|
105
|
–17.32
|
109
|
4.43
|
–14.17
|
* = compilation data from Govindaraju (1994) and GeoReM (georem.mpch-mainz.gwdg.de; 31 December 2012). A spreadsheet from Expedition 335 was also used as another quality control to check measurements. N = number of analyses. FeOT = all Fe as FeO. BDL = below detection limit, — = not applicable.