Table AT2. Pressure calibration of DVTPP, 7 June 2002. (See table notes.)

  PXDCR 88587 PXDCR 88579
Temperature coefficients
U0 (µs) 5.833194 5.878972
Y1 (°C/µs) –4036.649 –3969.749
Y2 (°C/µs) –12666.34 –11880.87
Y3 0 0
Pressure coefficients
C1 (psia) –68866.77 –66259.12
C2 (psia/µs) –694.2135 –2504.110
C3 (psia/µs2) 255224.2 223122.2
D1 0.029732 0.030630
D2 0 0
T1 (psia) 30.32667 30.23938
T2 (psia/µs) 1.025313 0.409012
T3 (psia/µs2) 62.84291 58.71868
T4 0 0
T5 0 0

Notes: For internal temperature sensor, temperature (°C) = Y1U + Y2U2 + Y3U3, U0 = temperature signal period (µs) at 25°C, U = temperature signal period – U0 (µs). For pressure transducer, pressure (psi) = C(1 – T02/T2)[1 – D(1 – T02/T2)], T = Pressure signal period (µs), U0 = temperature signal period (µs) at 25°C, C = C1 + C2U + C3U2, D = D1 + D2U, T0 = T1 + T2U + T3U2 + T4U3 + T5U4.