Table T2. Summary table detailing the sensor specifications, resolution, and accuracy for the Conductivity, Temperature, and Depth (CTD) YSI 6600 V2-4 sonde used during Expedition 325. (See table note.)

Sensor Range Resolution Accuracy
Conductivity 6560 sensor (mS/cm) 0–100 0.001–0.1 (range dependent) ±0.5% of reading + 0.00
Salinity (ppt) 0–70 0.01 ±1% of reading or 0.1, whichever is greater
Temperature 6560 sensor (°C) –5 to +60 0.01 ±0.15
Depth (m)      
Deep 0–200 0.001 ±0.3
Medium 0–61 0.001 ±0.12
Shallow 0–9.1 0.001 ±0.02
Vented level 0–9.1 0.001 ±0.018
Turbidity 6136 sensor (NTU) 0–1.000 0.1 ±2% of reading or 0.3, whichever is greater
Chlorophyll 6025 sensor (µg/L Chl) ~0–400 ~0.1 0.1
pH 6561 sensor (unit) 0–14 0.01 ±0.2
ROX optical dissolved oxygen % saturation 0–500 0.1 0–200: ±1 of reading or 1of air saturation, whichever is greater
BGA—Phycocyanin (cell/mL) ~0–200,000 ~160 1 cell/mL

Note: NTU = nephelometric turbidity units, Chl = chlorophyll, ROX = luminescent dissolved oxygen sensor, BGA = blue-green algae.