Table AT21. Temperature calibration of T2P. A. Calibration coefficients for data loggers and thermistors. B. Actual thermistor calibration data.
Table AT21A. Calibration coefficients. (See table notes.)
Calibration coefficient |
Logger Sn 2 | Logger Sn 4 |
---|---|---|
R1 | 1.000729 | 0.990307767 |
R2 | 176.136606 | –72.74909712 |
Calibration coefficient |
Thermistor 0509-1 |
Thermistor 0509-2 |
Thermistor 0509-3 |
Thermistor 0509-4 |
Thermistor 0509-5 |
Thermistor 0509-6 |
---|---|---|---|---|---|---|
A | 4.63276461E–04 | 5.43051013E–04 | 4.75887900E–04 | 4.82495572E–04 | 4.82688206E–04 | 4.71824656E–04 |
B | 2.10947147E–04 | 1.99981837E–04 | 2.08557612E–04 | 2.09282474E–04 | 2.08833095E–04 | 2.09263995E–04 |
C | 6.19690000E–08 | 8.35910000E–08 | 6.34430000E–08 | 6.20870000E–08 | 6.40080000E–08 | 6.27480000E–08 |
Notes: For temperature channel calibration (counts to ohms), Rt′ = R1 × [5/(x/65535 × 3.5/24900) – 24900] + R2, where x = counts and Rt′ = thermistor resistance in ohms. For thermistor calibration (ohms to Kelvin [Steinhart and Hart]), 1/t = A + B × Ln(Rt′) + C × Ln(Rt′)3, where t (°C) = K – 273.15 and Rt′ = thermistor resistance in ohms.