Fluid-sampling systems

During the two decades of CORK evolution described above, the corresponding fluid-sampling systems, sensors, and experiments also have evolved to improve our ability to collect pristine fluids from CORKs. The success of these systems requires that the hole recover from drilling disturbances, which include (1) surface seawater used as a drilling fluid to wash cuttings from the hole, (2) bottom seawater that enters naturally because a newly drilled hole is underpressured relative to neighboring crust, (3) drilling mud (clays to barite) used to keep the hole open, (4) cement used to seal casing, (5) grease used to lubricate steel threads, and (6) steel casing. The time required for borehole conditions to recover from drilling and installation disturbances varies but often requires many months to years. Given these variable conditions and the desire to detect borehole recovery events, CORK fluid samplers are designed to operate continuously under a range of conditions.

Fluid-sampling systems are categorized as either downhole or seafloor instruments. Seafloor instruments rely on a connection to the formation via umbilicals that span the length of the CORK and terminate at sampling bays on the wellhead. Seafloor fluid-collection systems have no size restriction and the advantage of being able to be recovered and modified on a yearly basis. The downside of seafloor sampling is that samples must ascend a tube before collection, and in situ temperature and pressure are not maintained. In contrast, downhole packages are size restricted and must remain in the borehole for years because opening the borehole disturbs the formation, especially if the formation is underpressured relative to hydrostatic. However, downhole packages are positioned in well-ventilated boreholes at in situ temperature and pressure. Fluids in this environment are at least as pristine as those that must ascend hundreds of meters of small-bore stainless steel or PTFE tubing to surface collection systems. However, recent results indicate that the stainless steel tubing does not contribute to the contamination of trace metals in solution provided the formation fluid is continuously ascending (Hulme and Wheat, 2010). Nevertheless, given the many obstacles related to CORK operations, redundant systems are critical in CORK design and sample-collection strategies. Below we describe in detail both types of fluid-sampling systems, starting with downhole systems.

Guiding principles for downhole systems

The principal goals of downhole fluid-sampling systems are to collect pristine fluids under in situ pressure, temperature, and chemical conditions and to conduct experiments using these fluids within a natural setting. Initially, there was a strong desire to distance the samplers from the steel casing, placing samplers in open boreholes whenever possible. This practice was used for several deployments until it was realized that many holes are unstable no mater how much hole is open below the samplers and how round (“to gauge”) the hole is, as determined with the downhole caliper tool. Hole instability is believed to be the reason why initial downhole packages from Holes 1025C, 1026B, and U1301B and the second deployment in Hole 1253A were not recoverable. Thus, beginning with Expedition 327 all of the downhole packages were safely deployed in perforated coated steel or fiberglass casing.

Placing samplers within the borehole at depth requires consideration of several critical constraints. First and foremost, all downhole samplers must fit within the inner diameter (ID) of the slick steel casing (tubing) used for CORKs. Typically, a 11.4 cm (4.5 inch) outer diameter (OD) pipe is used. This pipe is nested within a 27.3 cm (10.75 inch) OD and 25.4 cm (10 inch) ID casing. There have been many discussions about using a larger diameter casing for the CORK, which would provide more strength to the CORK and allow bigger packages to be deployed within the borehole; however, one restriction for using larger diameter casing is the diameter of the packers. For example, 11.4 cm (4.5 inch) OD casing has a 21.6 cm (8.5 inch) diameter packer, leaving a clearance of only 1.9 cm (0.75 inch) on either side of the packer. The use of larger tubing (5 or 5.5 inch OD) would require larger diameter packers, making the fit even tighter. Furthermore, the added strength afforded by the larger diameter casing is not significant (T. Pettigrew, pers. comm., 2010). Thus, downhole tools are restricted to a maximum of 8.9 cm (3.5 inch) in diameter, taking into account only one seal at the seafloor. In the case of multilevel CORKs, such as those intended for Expedition 336 with three distinct horizons, the deepest section requires that all samplers must have a diameter no larger than 6.3 cm (2.5 inch).

Another constraint for downhole instruments is that they must be able to withstand the temperature, pressure, and chemical conditions within the hole. At the sites currently instrumented, temperatures range from 3° to 65°C. Chemical conditions range from oxygenated seawater to highly altered fluids, from a pH of 7.8 to 12.2, from gas poor to gas rich and supersaturated at standard temperature and pressure (e.g., aboard ship), and from oxic to reducing with measurable amounts of dissolved sulfide, methane, and hydrogen. These conditions determine the suitability of materials for a particular deployment.

Other conditions are imposed by the nature of the CORK installation process. For example, every downhole package that has been recovered has been covered in pipe grease (typically a mixture of petroleum and metal [lead or zinc] powder). This grease is used generously between pipe joints during coring and CORK casing assembly. The grease is not restricted to the outside of downhole packages: it gets inside open cavities (e.g., Orcutt et al., 2011). In some cases, grease and sediment or rock fragments have been observed inside openings that act as cavities for drawing fluids into samplers. To date, no OsmoSamplers (1.2 mm ID) have been clogged with particles or grease from a borehole setting, but the presence of this grease should be taken into consideration for each experiment. For Expedition 336, nonmetallic and nonpetroleum-based lubricants are being considered.

Lastly, the sampling strategy and analytical program determine the length and number of samplers required. To date, only OsmoSamplers (described below) have been used to collect fluids in situ within boreholes. Typically, samplers are designed to draw ~0.5 mL of sample per day over the course of several years. This flow rate is appropriate for postcollection processing of the fluids to resolve samples on a day- to week-long temporal scale, and the volume is large enough to measure chlorinity (via titration; requiring 0.05 mL), major ions (via inductively coupled plasma–atomic emission spectroscopy [ICP-AES]; 0.05 mL; Ca, Mg, Na, K, Sr, and S), minor ions (ICP-AES; 0.25 mL; Sr, Li, B, Mn, Fe, and Si), and trace elements in seawater (inductively coupled plasma–mass spectroscopy [ICP-MS]; 0.1 mL; V, Mo, Cs, Ba, Rb, Y, and U). If a more ambitious sampling plan is desired, then more fluid can be collected or one could combine adjacent samples to include isotopic measurements for O, H, S, Li, and B. The length of the deployment and desired sample rate result in sample tubing that is kilometers in length, and this can take up a significant length of vertical space inside the CORK casing, even with the tubing tightly coiled. The combination of osmotic pump and sample coils for a single typical downhole OsmoSampler package is ~3–6 m in length.

We again stress here that instruments that are placed within the borehole should be deployed with the intention of leaving them for at least 3 y and preferably 4–5 y. Such a long time span is required because each time a hole is drilled a considerable amount of seawater is introduced into the formation during the drilling process. The time required for the hole to return to in situ thermal and chemical conditions can take weeks to years (e.g., Wheat et al., 2003a, 2010b; Solomon et al., 2009). Even after in situ conditions return within the borehole, if the borehole is underpressured, seawater will flow into the borehole once the seals are removed during recovery of downhole instruments and the borehole must again recover; however, recovery is much quicker because the shear volume of seawater entering the hole is diminished relative to that introduced during drilling.

No fluid samplers or sensors that require electrical connection with the seafloor have been deployed in the borehole (only thermistor cables) because of the complexity of such a deployment, the lack of investigators wishing to modify their instruments to fit within the borehole, and the added expense of a cable. Complexities in deployment include last-minute potting of electrical bulkheads because the length of electrical cables cannot be determined until the hole is drilled and cleaned just prior to CORK deployment, and at least one connection must be made at sea. Bending a predesigned cable has not worked (i.e., bent thermistor cables usually result in poor or no data from the section below the bend). Another issue with deploying instruments that require an electrical connection from the drillship is that any data logger or unit used to access the electrical cable within the borehole must fit within the 11.4 cm (4.5 inch) diameter casing (10.2 cm ID) or be attached to the wellhead. If the unit is not attached or buoyant, design constraints must account for it falling off the top of the well and disrupting other instruments and valves on the wellhead. A reasonable approach is to deploy a connector within a buoyant housing that can later be handled using a submersible.

In conclusion, the most significant issues with downhole fluid samplers and sensors are the confined space, consideration of appropriate materials, temporal length of the experiments, and difficulty and cost of providing communication and power.

Downhole OsmoSampler

Design specifications

The critical parameters for OsmoSampler design are outlined above. These parameters determine the number of membranes, the structural integrity of the packages, the types of sample coils, and the alignment of pumps and sample coils needed to achieve a variety of applications. We discuss each of these parameters below, with the caveat that the discussion is limited to what has been accomplished to date. There are many other possibilities, limited only by one’s imagination.

OsmoSamplers use the osmotic pressure that is created across a semipermeable membrane by solutions of different salinity to continuously draw sample through a small bore (1.2 mm ID) tubing (reviewed in Jannasch et al., 2004). This process is known as forward osmosis (FO) because it involves pumping a solution with the osmotic pressure gradient, rather than in opposition to it, as with a reverse-osmosis (RO) system (Cath et al., 2006; Wang et al., 2010) (Fig. F7; see also OSMO in CORK in “Supplementary material”). RO systems are used in traditional desalinization plants, where salt water is forced at high pressure through semipermeable membranes, effectively filtering solutes from seawater. The FO pumps that have been used in OsmoSamplers to date are based on a small-scale design that was initially developed for medical research purposes to deliver drugs to laboratory animals at a constant rate. The general equation for the flux of water (J) across a semipermeable membrane using a single osmotic driving agent (solute) is

J = K · A · (σΔπ – ΔP),

where water flow (J) is directly proportional to permeability (K) and the area (A) of the membrane, σ is the salt reflection coefficient and represents the ability of the membrane to reject the solute (approaches unity with an ideal semipermeable material), Δπ is the osmotic pressure differential driven by the difference in solute concentrations across the membrane as well as the type of solute used, and ΔP is the hydrostatic pressure differential between the intake and outflow of the system (Theeuwes and Yum, 1976). For an ideal FO system with negligible hydrostatic pressure, zero solute concentration in the feed (i.e., distilled water) solution, and an ideal semipermeable membrane, the flux of water through the system is governed only by the physical properties of the membrane and the osmotic pressure of the solute:

J = K · A · π.

Thus, osmotic pressure is used to draw fluids into the small-bore tubing that is connected to the distilled water reservoir of the osmotic pumps. Small-bore tubing is used because the combination of the small bore and slow flow rate minimizes dispersion, resulting in pluglike flow (Jannasch et al., 2004). Upon completion of the deployment, the sample coil is disconnected from the pump. Lengths of tubing (typically ~1 m) are cut, and the fluids are discharged out of these discrete sections into sample vials; a continuous sample collection device cannot be used because dispersion within the tubing during extraction will “homogenize” the sample and decrease temporal resolution. Time stamps for the individual samples are calculated given the start and end dates and number of discrete samples (assuming constant lengths/volumes of tubing cut), provided the temperature is constant for the deployment. All of the other variables are constrained through the pump and membrane design.

Membranes and pumps

At the crux of all osmotic pumps (Fig. F7) is the membrane that allows water to flow through it, yet prevents the transfer of ions. Many materials can be used for a membrane. These materials can be commonly found in a local hardware store where RO products are sold. However, the key to a good membrane material is longevity and reproducibility at depth and temperature.

We have had excellent success with cellulose acetate (CA) membranes sold by Alzet. These membranes are rigid and commonly used in pharmaceutical studies for drug-delivery mechanisms. Thus, these membranes are subject to strict quality control measures to ensure a uniform product. This product also has a well-defined response to temperature and is not influenced by oceanic pressure. Although these membranes come in several materials, they are available in two sizes. In contrast, RO thin-film membranes come in a variety of sizes, are paperlike in texture, and can be acquired in rolls or sheets from local water agencies. RO membranes are not as uniform as the Alzet membranes, but some types of RO membranes pump orders of magnitude faster for an identical surface area. The reason for the reduced pump rate of the Alzet membrane is that the thickness of the semipermeable (active) layer is greatly increased. In addition to greatly reducing the pump rate, it makes the membrane self-supporting. In contrast, thin-film RO membranes experience solvent fluxing into the porous support layer (often paper) and dilute the solute with time, reducing the osmotic driving force and the pump rate. This effect is called internal concentration polarization (Gray et al., 2006). The decision to use either RO or Alzet membranes is typically dictated by the volume of fluid needed versus the length of the deployment, with multiyear samplers necessitating the use of the nonvariable Alzet membranes.

An osmotic pump consists of a membrane that separates a freshwater reservoir from a salty reservoir (Jannasch et al., 2004; Wheat et al., 2000). As described above, the rate of water transferred across the membrane is a function of the chemical gradient, temperature, pressure, type of membrane material, and surface area of the membrane. Osmotic pumps rely on a chemical gradient that is defined by distilled water on one side of the membrane and saturated salt on the other side. Noniodized table salt is used on the saturated salt side because it has a high solubility (greater gradient) and is inexpensive, easy to purchase, and nontoxic. Noniodized salt is preferred because it appears to pour more easily. If one were to use a different salt with a lower solubility, the pump rate would be less, given all of the same parameters above. Thus, to maintain a uniform pump rate during a deployment, the sample coil is filled with distilled water and enough salt is added to the reservoir opposite the distilled water such that at the end of the deployment some crystalline salt remains. It is important to make sure that Alzet membranes are completely filled with salt and that osmotic pumps are deployed vertically to keep salt crystals against the membrane (i.e., a tilted pump may allow the salt to settle away from the membrane, which will affect the ionic gradient if a freshwater lens forms in the pump). Increases in temperature usually increase the rate of pumping, as diffusion is quicker at higher temperatures.

Lastly, pressure can be a factor with some pump systems. An increase in pressure sometimes increases the pump rate of RO systems. We believe this results from the removal of air from spaces within the RO membrane because upon return to the sea surface these membranes continue to pump at the in situ rates. Typically, we have designed OsmoSamplers with Alzet membranes because there is not a strong demand for >0.5 mL of sample per day and enough membranes can fit within the allotted diameter of the pump to get the desired pump rate for a given temperature. Typically, pumps use 12 Alzet 2ML1 membranes for deployments at seafloor temperatures, providing ~0.7 mL of sample per day, or 250 mL per year (this volume fits easily within one 305 m long spool of 1.19 mm ID tubing).

In contrast, some RO membranes (47 mm diameter) pump fluids at a rate of 50 mL per day. The primary problem with RO membranes is the internal concentration polarization effect, which is caused by a buildup of solute in the support layer over time, reducing the osmotic pressure gradient and the pump rate (Gray et al., 2006). Thus, determining time stamps for individual samples is less certain. There is currently only one commercially available FO-specific membrane, produced by Hydration Technologies Inc., and it has performed well when used in FO systems (Cath et al., 2006). Wang et al. (2010) conducted experiments with an alternative to the commercial membranes by synthesizing double-skinned CA membranes that minimize the internal concentration polarization by preventing solute from entering the support layer. We have not, as yet, applied thin-film FO-specific membranes to OsmoSampler products, but we are currently experimenting with thin-film designs using RO and FO membranes.

Another key issue with osmotic pumps is the mechanism that seals them from the outside, except for the intake where sample tubing is attached and the outflow where saturated salt water exits. To make such sealed environments, a combination of O-rings and two-part epoxy has been used. If O-rings are used, two are suggested, with one serving as a backup. A potential problem with two O-rings is the presence of air between the two O-rings. Constructing the pumps under water in a large tub or garbage can will eliminate this concern. The Expedition 336 pumps will use only one O-ring, avoiding underwater construction, but the O-ring has a substantial diameter to maximize the sealing surface. Epoxies are effective, especially for holding Alzet membranes in the pump housing. For deployments at <40°C a two-part polyurethane compound (Fluid Polymers; HMP-60) has had great success. However, at higher temperature (certainly at 64°C), this epoxy becomes soft and pliable and flows. Hysol ES1902 is a suitable two-part epoxy for warm (>50°–80°C) environments. Early samplers relied solely on epoxy, but recent deployments include an O-ring and epoxy design to provide a redundant sealing mechanism for Alzet membranes. In contrast, RO membranes, because of their paperlike consistency, are held in place with an O-ring seal with support grids on both sides of the membrane.

One must also consider what material to use to hold and protect the membranes while maintaining a leak-free system. Polycarbonate and translucent polyvinyl chloride (PVC) are typically used for this purpose because both materials are easy to machine and epoxy readily seals to both. Polycarbonate is used for cool (<40°C) oxygenated or mildly reducing environments. However, polycarbonate should not be used in reducing environments with hydrogen sulfide. Hydrogen sulfide impregnates polycarbonate in seawater, making it brittle. Once retrieved, hydrogen sulfide will degas and the polycarbonate will return to its initial state, flexible but clouded. Therefore, PVC is used for reducing environments. Clear PVC tubing is not readily available in many sizes; however, individual production runs are available. Clear PVC allows one to observe conditions within the pump and track the status of O-rings during assembly.

As with the membranes, the pump assembly can be held together with O-rings and machined grooves or a combination of O-rings and epoxy. If epoxy is used, the pump cannot be modified or refurbished. One also must be concerned with getting a complete seal with the epoxy and ensuring the epoxy is flexible enough to withstand the potential beating that OsmoSampler packages take while being deployed. Many pumps have been deployed using only epoxy without incident, but a double O-ring seal is suggested for deployments at temperatures of >40°C.

Lastly, we do everything possible to protect the membrane from ambient water (e.g., adding extra spools of sample tubing). However, there are times when one cannot retrieve the sampler on time and sample overflows from the coiled tubing to the freshwater reservoir. It is always wise to check the salinity of this reservoir following recovery to determine whether it has been compromised. If seawater reaches the distilled water reservoir, salt from the fluid being sampled increases in the freshwater reservoir because only water passes through the semipermeable membrane. The presence of salts in the freshwater reservoir decreases the salt gradient, slowing the pump rate. Furthermore, ambient fluids contain microbes and dissolved chemical species that can be detrimental to membrane integrity. Such was the case for one deployment in ODP Hole 1026B. Borehole fluids (and potentially microbes) dissolved several membranes completely, stopping the pump.

Strength elements and structure

Considering the demands of deployment and recovery operations, downhole packages must be well protected and have a strength member to provide support. Several options have been tried. Initially, OsmoSampler packages were add-ons and were designed to be compatible with the existing CORK design and modes of deployment and recovery. To maintain continuity of the thermistor cable, initial OsmoSampler packages were attached to the cable. In some cases, OsmoSampler packages were attached mid-cable. These noncentered deployments were potentially problematic. Other packages were designed so that the cable could slide down the middle of individual OsmoSamplers. As a result, OsmoSampler packages were placed in the middle of the thermistor cable in ODP Hole 949C and near the bottom of the thermistor cable in ODP Holes 1024C, 1025C, 1026B, 1027C, and 1200C (Shipley, Ogawa, Blum, et al., 1995; Davis, Fisher, Firth, et al., 1997; Salisbury, Shinohara, Richter, et al., 2002). In each of these cases the strength member consisted of a stainless steel rod or tube at the center and a PVC or polycarbonate tube on the outside for protection.

For added strength, OsmoSamplers were placed in steel pipes during Leg 205, with their intakes extending from the ends of the pipe; however, several issues arise with steel tubing. First, steel oxidizes, coating the samplers with rust and making it more difficult to obtain a “clean” sample when removing and cutting up sample coils. Second, few shops can machine long, straight pipes, and transport requires cranes and forklifts.

Given the difficulties with steel tubing, we suggest using a 1.27 cm (0.5 inch) diameter stainless steel rod as the strength member. This rod may include a groove that allows small-bore tubing to be placed and protected so that it can connect to another part of the package or discharge into a desired area. OsmoSamplers fit around this strength member with a snug-fitting outer protective shell of polycarbonate or clear PVC. Galvanized shackles are then connected to the stainless steel support rods to connect to other OsmoSampler packages, plugs, Spectra cable, or sinker bars. Shackles will be discontinued during Expedition 336 because 1.25 cm thick shackles (required for strength) are too large for small-borehole conditions. Instead, a novel pin-locking system has been developed and will be deployed. OsmoSampler components were assembled and placed within a single PVC or polycarbonate housing for protection. In some cases, the outer diameter of the pump was designed as the outer protection; thus, assembly is modular, with separate protective housings for different parts of the sampler (Fig. F8).

Sample coils

OsmoSamplers rely on small-bore tubing for nonturbulent (dispersive), pluglike flow (Fig. F7). Two sizes of tubing have been used: 0.8 and 1.19 mm ID. The larger diameter is more common because it holds more water per unit length (0.50 and 1.11 mL, respectively, for a 1 m long section). The larger diameter tubing has been purchased in bulk from Professional Plastics (Elosia 800-338-2011). Tubing is purchased as a continuous roll that is 305 m (1000 ft) long with a 1.19 mm (0.047 inch) ID holding 339 mL. The OD for fluorinated ethylene propylene (FEP) tubing is 2.0 mm (0.079 inch), and for copper it is 1.8 mm (0.071 inch). The purchased material is spooled on PVC spools with a lathe or any variable-speed motor. These spools slide over the internal stainless steel strength rod and fit within the protective outer tubing. It takes ~2 h to spool a 305 m long coil by hand or 30 min on a lathe.

Before deployment, all sample spools are filled with distilled water. A peristaltic pump is used to inject water into the coil to make sure there are no leaks, and then 10% hydrochloric acid is pumped into the coil to remove any trace metal contaminants. This acid remains in the tubing for ~1 week. Longer storage is problematic because air bubbles will form in the tubing, which drastically increases the resistance when flushing the coil, requiring a much slower pump rate. Distilled water is pumped into the coil to remove the dilute acid ~1 day before assembly. Coils filled with distilled water are stored at a constant temperature to avoid the production of air bubbles. Filling the coils with degassed distilled water can alleviate this concern if the coils are deployed within days. Air bubbles greatly restrict flow and are an issue for shallow deployments. For deep deployments, air bubbles will dissolve at depth, but the volume of these air bubbles is replaced with seawater at the intake. This seawater influx, if substantial, could affect the calculated sample time stamps. Copper coils are filled with distilled water to check for leaks and are then flushed with distilled water before assembly.

Any number of coils can be used, depending on the analytical program and duration of the deployment. Coils are attached with “Upchurch” flangeless compression fittings (¼-28 threaded polyether ether ketone [PEEK], or Delrin, nuts with a inch ID) using readily available nuts, ferrules, and unions. Such fittings must be tight, but not so tight as to pinch off the tubing at the ferrule. Practice with a connection and tighten by hand as much as possible. Release the connection and check to verify that there is not a significant constriction. Furthermore, it is important to check the tubing for any kinks or drastic bends that may crimp the tubing. We tighten fittings as tight as possible and then push water with a syringe through the connected coils. This confirms that the union and fittings are not leaking and that there is no significant constriction. Seven 305 m long coils have been attached to a single pump for a 5–6 y deployment. The connection to the pump is always supported with PTFE tape to ensure that the nut will not unscrew at depth. Likewise, all fittings at the pump are wrapped in PTFE tape to minimize the potential for leaks because different materials have different thermal expansion coefficients. PTFE tape can be used on any connection.

Geochemical, microbial, and hydrologic configurations

The combination of pumps with different numbers and types of membranes and sample coils provides for a range of possible OsmoSampler packages (e.g., Fig. F2; see also OSMO in CORK in “Supplementary material”). The standard package contains one pump and a number of PTFE coils. This type of sampler is ideal for ship- and shore-based analysis of the major and minor ions in seawater and, more recently, for microbiological characteristics. Unfortunately, PTFE is permeable to gas exchange with the formation fluid. Thus, if the gas concentration (e.g., carbon dioxide, oxygen) changes during the deployment, only the most recent composition will be retained.

The gas package is ideal for maintaining concentrations of dissolved gases. This package uses copper tubing instead of PTFE. Similar to the standard package, the gas package uses only one pump with a number of copper coils connected by fittings and unions. One issue with such small-bore copper tubing is the ability to make a clean cut. Dental tools are ideal for opening the tubing after an imperfect cut. Following recovery, the ends of the coils are crimp sealed for shore-based analysis if analysis is not possible at sea. Freezing the crimped coils can keep the solution from mixing; the expansion of water upon freezing does not break this small-bore copper tubing. Note that highly gas saturated environments may lead to an overpressure in the OsmoSampler sample tubing, which can cause sample smearing or, in the worst case, complete loss upon recovery. The use of valves to close OsmoSampler coils prior to recovery can maintain in situ pressures and gas concentrations (e.g., Lapham et al., 2008) and were included in the downhole package deployed in Hole 948D.

Another class of OsmoSampler includes the acid-addition and BioOsmoSampling System (BOSS; Girguis et al., 2008) packages. Both packages comprise (from bottom to top) a sample coil, a lower pump, a second coil, a T-connector, a series of sample coils, and a second pump. The lower pump pulls fluids into the lower sample coil, acting as a standard OsmoSampler, and the saturated salt fluid is expelled into the second coil to disperse the coil contents. In the acid-addition OsmoSampler package, this second coil is filled with diluted, subboiled hydrochloric or nitric acid, often including a tracer (e.g., a rare earth element) to document the extent of dilution. Thus, the lower pump pushes the acid out and into the T-connector, which has one branch open to the borehole (sample intake) and the other branch attached to the upper series of sample coils. These coils are attached to a second pump that pumps at a rate that is at least four times faster than the lower pump. Thus, coils above the T-connector capture borehole fluids from the T-connector’s sample intake as well as acid expelled by the lower pump, acidifying the sample in situ. This preserves the sample for trace metal analysis and prevents microbial oxidation and mineralization within the sample tubing. This process also may dissolve small, suspended particles that are incorporated into the sample stream. The BOSS package is configured identically to the acid-addition package, except a biocide (e.g., ammonium sulfate, glycerol, or mercuric chloride) is used in the second coil instead of dilute acid, preserving samples for microbiological analysis.

The Flow-through Osmo Colonization System (FLOCS) (Orcutt et al., 2010) package is designed to use an osmotically driven pump to slowly pull formation fluid into microbial colonization devices for in situ experimentation. The term “FLOCS” refers to a series of chambers filled with various minerals and other substrates for examining microbial growth under controlled conditions. These flow-through chambers are connected to the intake of PTFE sampling coils and an osmotic pump. Chambers are designed with minimal excess pore space so that borehole fluids flow through the chambers and into the sample coils. These coils allow one to characterize reactions within the chambers by comparing results to those from a standard package.

The enrichment package is a combination of the FLOCS and acid-addition packages. This package is essentially the same as the acid-addition package, except the FLOCS experiment is positioned between the T-connector and the upper series of sample coils. The coil above the first pump is filled with enrichment material instead of dilute acid. The enrichment solution could contain any number of substances, including isotopically labeled carbon, amino acids, nutrients, sugars, or any rate-effecting substance that will mix with borehole fluids in the FLOCS chamber (Fisher, Wheat, et al.).

A tracer package is used to introduce a chemical tracer into the borehole environment at a constant rate to examine fluid mixing within the borehole. This package consists (from bottom to top) of a series of sample coils, a single pump, and a similar number of additional sample coils. Basically, the tracer package is a standard package with coils attached to the salt water outflow of the pump. Like the acid-addition package, saturated salt water is pumped into these upper coils, expelling the fluid, which contains some type of tracer. A variety of tracers can be used; however, the tracer solution should be many orders of magnitude more concentrated than what occurs naturally within the formation. Historical deployments used Cs, Rb, and rare earth elements (Fisher et al., 2005). Other potential tracers include but are not limited to SF6, labeled isotopes, and fluorescent microspheres.

Lastly, a flowmeter package has been developed for documenting fluid flow speeds and direction within a borehole (Jannasch et al., 2003; Solomon et al., 2009). This package uses a series of three standard packages and one tracer package. Four intakes are positioned within a machined titanium solid stock with two perpendicular crossing holes. Intakes are positioned near the ends of these holes, with the tracer injected in the middle where the two holes intersect. Thus, the presence and concentration of the tracer at the various intakes is a function of the direction and rate of flow through the system.

Each of these OsmoSampler packages includes space for a self-recording temperature sensor. Because of the pump design, the only factor determining flow rate is temperature. Two different temperature units have been used at full ocean depth. The Antares logger (Titanium 1857C MTL, 16 byte) has a resolution of 0.002°C and can be purchased with a low (0°–45°C) or high (>40°C) temperature range and with an extra battery for a 5 y deployment. Onset (U12-015 titanium, 12 byte) also sells a temperature unit with a 5 y lifetime, but the resolution is 0.1°C at lower temperatures (<30°C) and 0.3°C at ~60°C (Fisher, Wheat, et al.).

Weights, rope, and plugs

A complete downhole string generally comprises (from bottom to top) a sinker bar, a number of OsmoSampler packages, a plug, Spectra cable, possibly a middle sinker bar, and a top plug. The weight of the bottom sinker bar deployed from the drillship is typically 91 kg (200 lb) in seawater. Together with the weight of the OsmoSampler packages, plugs, and middle sinker bar, the weight of the string is ~200 kg in seawater (400–500 lb). The sinker bar not only acts as a useful guide to position the rest of the string in the hole, its weight also allows one to detect when the string is released from the wireline mechanism during deployment. Because of the weight of the ship’s wire, 200 kg is about the limit for detecting the release of the string in good weather (minor heave). For submersible-aided deployments, the bottom sinker bar weighs less (68 kg [150 lb]) because of handling issues on deck and on the seafloor. In contrast, middle sinker bars do not weigh as much (5–50 kg), serve to help pull down the Spectra cable, and are typically 5 cm in diameter.

Downhole plugs are generally made of stainless steel and seal with an O-ring surface (Fig. F8). Because of the confines of the well and the desire for multiple levels, most seal surfaces can accommodate only one O-ring. Downhole plugs are generally 10 kg in water. The weight of the plug and the materials hanging below need to exceed the buoyancy force in crust that is overpressured relative to hydrostatic. On ridge flanks this could require several hundred kilograms. In subduction zones this could require even more weight or the necessity of a latch-type plug at the seafloor. Thus, depending on the environment, the surface latch may be a simple O-ring gravity seal or an O-ring latch seal. The only concern with a latch is the potential for failure due to corrosion and the inability to access the borehole. Several configurations have been used for top plugs, each requiring different tools for retrieval (see “CORK-IIs and L-CORKs”).

Spectra cable is made from Spectra Fiber 900/1000/1200 (Puget Sound Rope), an ultrahigh–molecular weight polyethylene fiber. Honeywell Spectra is used because it is buoyant (or at least neutrally buoyant) in seawater, strong, and durable in a variety of chemical and thermal conditions. It withstands temperatures of 65°C in reducing conditions and fluids with a pH in excess of 12. Spectra cable (0.95 cm [⅜ inch], with a minimal tensile strength of 6,300 kg [13,900 lb]) is carefully measured by the manufacturer; however, the rope stretches with weight. Once the rope is stretched under load (>200 kg, which is the weight of a typical string during deployment), it maintains this new length, ~2% longer than the length purchased from the manufacturer.

Guiding principles for seafloor systems

Seafloor samplers and sensors do not have the myriad restrictions and concerns that must be considered in the design of downhole systems. For example, since these instruments are bathed with cold seawater, corrosion and gas exchange issues are minimal. Also, there is no space restriction at the seafloor, allowing large instruments to be deployed with a submersible or on an elevator. An elevator is a platform that allows many or heavy items to be placed on the seafloor. Elevators are released from the ship near the target location. After they descend, they are located by submersible and moved into position near the CORK. Tools, samplers, and instruments from the elevator are used and replaced. When work is completed the elevator’s drop weights are released, allowing it to float to the surface. Thus large and complex instruments can and have been deployed to sample fluids that discharge naturally from overpressured holes in permeable crust and those that are extracted through umbilicals that bisect a hydrologic horizon of interest. The type, number, and size of umbilicals have differed through the years as interests and goals have changed. Similarly, valve and mating systems for connecting instruments to umbilicals have changed. Below, we describe the different umbilicals and valves and the evolution of both. Then we describe a number of sensors and samplers that have been attached to these valves and some that have been lowered into boreholes using underwater vehicles.

Seafloor systems

Umbilicals

Umbilicals are lengths of tubing that terminate at the seafloor and at depth, allowing fluids to be transported from the subseafloor to the seafloor. An umbilical can consist of a single length of tubing or a collection of tubing bundled in a hard plastic sheath for protection. The initial geochemical tubing (umbilical) was thin-walled PTFE tubing (Davis, Mottl, Fisher, et al., 1992). The umbilical was designed to act as a conduit for fluids to flow from depth and ascend without contacting the mild steel casing, minimizing artifacts and providing a pristine formation fluid sample from the hydrologic horizon of interest. However, the wall of the PTFE tubing was too thin; the tubing collapsed when it was attached to the thermistor cable, and it was not robust enough to withstand handling on the rig floor during deployment.

Hardened armored tubing has been used on all CORKs for geochemical analysis since Leg 205. At those sites where the hydrologic zone of interest ends in sediment (e.g., Hole 1255A), umbilicals terminate in miniscreens placed within a carbolite pack, which prevents sediment particles from clogging the screens. Carbolite packs are permeable (~2 × 10–10 m2) and are made from relatively nonreactive ceramic material (aluminum oxide ceramic with a grain size of 400–600 mm and a porosity of ~30%) (Mikada, Becker, Moore, Klaus, et al., 2002). This material is strapped around a screened portion of the casing, allowing water to flow from a larger area into the carbolite pack and finally into the miniscreen. An inflatable packer is then used to seal off the hydrologic horizon from the remainder of the hole. This design allows for multiple horizons to be sampled, provided packers and sediment seal the hole. This CORK design also leaves the borehole open for instruments.

A variety of tubing sizes and types have been used on CORK-IIs to access seafloor horizons for fluid sampling. Most of this tubing is made of stainless steel, and the diameters are chosen depending on the hole and desired sampling program. The first consideration in umbilical design is to assess whether the formation is overpressured or underpressured relative to hydrostatic. The former produces a free-flowing system, provided that the formation is permeable. For this situation, larger diameter tubing is desired (1.27 cm [0.5 inch]), minimizing the impedance of natural flow. The latter requires a pump to draw fluids up the tubing. If a mechanical pump is chosen to extract fluids from depth, then large (1.27 cm) tubing is desirable to reduce frictional drag during mechanical pumping (e.g., backpressure). However, small-bore tubing is desirable if the intent is to collect fluids with an OsmoSampler at the surface. Stainless steel tubing with 0.32 cm (0.125 inch) OD and 0.175 cm (0.069 inch) ID was used for an umbilical deployed during Expedition 301 to minimize the volume of fluid in the tubing, allowing osmotic pumps to recover fluids in a reasonable amount of time.

The observed effect (i.e., concentration) of extended exposure of borehole fluids to mild steel casing on fluid composition (e.g., iron, H2, H2S), and potentially on microbial community compositions, resulted in the use of an armored tubing with an ethylene tetrafluoroethylene (ETFE) fluoropolymer wetting surface. Armored ETFE was chosen because it is nonreactive. It was first used in IODP Hole U1301B (1.2 cm OD). To minimize contamination, titanium tubing connections, valves, packer pass-throughs, and miniscreens were used. Unfortunately, this line failed to produce significant flow, despite substantial mechanical pumping, possibly because of damage to the umbilical during deployment. New CORK installations during Expeditions 327 and 336 are and will be equipped with multiple stainless steel and ETFE umbilicals to service a variety of fluid-sampling needs and provide redundant systems.

The desired configuration for Expedition 336 is a single 1.27 cm (0.5 inch) diameter armored ETFE tubing and two stainless steel tubes (0.32 cm [0.125 inch] and 0.634 cm [0.25 inch]) for each depth horizon. The smaller stainless steel tubes are desired because the boreholes are believed to be underpressured relative to hydrostatic. Thus, when the sample valve is opened, seawater will flow naturally into the tubing and descend to the formation. Thin-film OsmoSamplers (see “Design specifications”) will be attached to the small-diameter tubes, minimizing the residence time that fluids spend within the umbilical tubing. The larger bore ETFE tubing is designed for chemical and microbiological experiments that require large volumes of fluids at discrete sample intervals. This larger tubing greatly minimizes frictional drag and the residence time that fluids spend within the tubing relative to a slower osmotic pump.

Valves and bays

Originally, the valves and tubing connecting seafloor samplers and sensors to the horizon of interest consisted of a ball valve with a male Aeroquip quick-disconnect connector (complete assembly: number FD72-1000-08-10; female half: number FD72-1001-08-10; male half: number FD76-1002-08-10) (Figs. F9, F10; see HANDLE in CORK in “Supplementary material”). However, Aeroquip valves are problematic; they are often difficult to connect, and without making in situ real-time measurements (e.g., temperature and oxygen) one is never sure that the system is fully coupled. Even today, discussion continues regarding the quality of data and the potential for problems with this connection.

As issues related to the Aeroquip connector grew, CORKs deployed during Expedition 301 included a series of valves and a unique sample distribution system (Fig. F4; see GUIDE in CORK in “Supplementary material”). This system accepts interchangeable plates upon which samplers can “sip” (for overpressured systems) or pump fluids from the formation. Some of the stainless steel umbilical tubing is 0.32 cm (⅛ inch) OD with an ID of 0.175 cm (0.069 inch), allowing seafloor OsmoSamplers to recover fluid within a reasonable period of time. OsmoSamplers and other samplers are attached to the wellhead using a metal plate that guides sample intakes over titanium fittings on the wellhead (Fisher et al., 2005).

The plate system of sample access is sometimes difficult to deploy, and successful deployment is dependent on the pilot. Because of this difficulty and those caused by using the Aeroquip connector, a new sample connector was developed for Expedition 327. This connector relies on a double O-ring seal. The O-rings reside in an easy-to-manipulate handle that is in position only when the O-rings are clearly sealed in the base (Fig. F5; see HANDLE327 in CORK in “Supplementary material”). The base, made from ultrahigh-molecular weight polyethylene, houses the valve and the sample port. This housing guides the sample handle to and over the 1.27 cm (0.5 inch) diameter PEEK tube, which is an extension of the umbilical. This configuration also removes the necessity of the large plates required for Expedition 301 CORKs, allowing flexibility in package design (e.g., deployment of samplers and experiments in a milk crate), which can be hung from a series of hooks below the valves or from cut-outs in the bulkhead gussets. These new connections were first deployed on CORKs deployed in Holes U1362A and U1362B during Expedition 327. The new connections will also be used during Expedition 336, but the PEEK ID will be identical to the umbilical ID and the void space in the handle will likewise maintain the same ID.

With the addition of umbilicals for geochemical and microbial sampling, the CORK wellhead was modified to fit extra valves and make room for sampling systems. The wellhead was transformed during Leg 205 to have partitions (bays) around the central strength member. During Expedition 301 the partitions were designated for pressure, geochemistry, and microbiology sampling, but any bay can be used for any type of scientific measurement or collection. This partitioning allowed different groups to design sampling capabilities without affecting others.

Presently the bays are 198 cm in length, and they will be extended during Expedition 336 another 60 cm to accommodate new samplers that will be deployed from the drillship. Samplers must fit within a radius of ~76 cm extending from the 11.4 cm OD inner mandrel (strength member) to accommodate the diameter of the wireline camera system on the JOIDES Resolution. No outward radial restrictions apply once the wellhead is on the seafloor.

Seafloor sampling and sensor systems

Sampling and sensor systems on the seafloor rely on the umbilical and valve systems to obtain borehole fluids from specific depth horizons. If the horizon at the end of the umbilical is permeable and overpressured relative to hydrostatic, fluid will naturally flow up the tubing when the valves at the seafloor are opened. In contrast, if the horizon is underpressured, a pump (mechanical or osmotic) is required to pull formation fluid up through the umbilical tubing to a sampler on the seafloor. For example, formation fluids have been flowing from the décollement, ascending the umbilical in ODP Hole 808I, and discharging at the seafloor for the past several years. Thus, all that is needed to collect these fluids is a sampler with a small snorkel (inlet).

An example of a seafloor collection device that tapped into an umbilical or wellhead valve is the “BioColumn,” a series of microbiological and geochemical samplers designed to take advantage of the free-flowing system in ODP Hole 1026B (Cowen et al., 2003) (Fig. F11A). The BioColumn channeled borehole fluids past temperature and flow sensors, through particle filters (>0.4 μm pore size), and into an organic compound-scavenging resin column to collect samples for fluid composition, biomass, DNA, and lipid analysis. BioColumns were deployed for days to 1 y, with remarkably constant, unassisted flow rates of ~4.4–5.4 L/min (Cowen et al., 2003; Cowen, 2004).

However, the greatest disadvantage of the initial BioColumn is that the fluids collected from the top of the borehole must flow past 295 m of mild steel casing. The extent to which the fluids are altered by interaction with casing, cement, and other drilling remnants has been documented by sampling fluids that vented from the borehole after the data logger was removed and the borehole was well ventilated (Wheat et al., 2004). Fluids that flowed unrestricted through the steel casing and out the 10.6 cm diameter opening at the top of the CORK into seawater were contaminated with high dissolved and particulate concentrations of several metals, including iron (Wheat et al., 2004). This leads one to wonder if the surface of the iron liner could support a biofilm community that could contribute to the cellular biomass and molecular diversity observed in the ODP Hole 1026B fluid and BioColumn samples, considering that the BioColumn restricts flow to ~4.8 L/min, with a maximum temperature of 59°C at the CORK valve (Cowen et al., 2003).

Submersible positive displacement pumps were developed to draw fluids from subbottom horizons using the umbilical system at rates sufficient to deliver tens of liters of fluid within one to several hours. Initially, pumps were downstream of the samplers, which limited sampling and plumbing options and made the pumps more susceptible to contamination from ambient bottom seawater. Subsequent use of a PTFE and titanium pump head that is magnetically coupled to the motor drive improved performance and increased versatility (i.e., pump draws fluid up the CORK’s umbilical and then pushes fluid through or past sensors and samplers). Pumps are connected to the CORK fluid delivery line with an Aeroquip connector on CORKs deployed prior to Expedition 327 and with the new O-ring-based connector design on CORKs deployed during Expedition 327 or later. It is essential to evaluate the seal in situ, if possible, using available information such as temperature, current to the pump, and the pump’s programmed and actual revolutions per minute.

These fluid pumping systems have been used effectively with both in-line samplers (e.g., GeoMICROBE sled) and end-of-line samplers. The latter include in situ filtrations and the filling of large-volume bag samplers (e.g., Wommack et al., 2004) (Fig. F11B). Such large-volume in situ filtration systems have an autonomous mode that can be connected to the wellhead for several days of continuous operation. Another mode places the pump on the submersible. The sampler can be transported to the seafloor on an elevator or on the submersible. Alternatively, pump systems can have their pumped effluent directed to an exit port that can be accessed by a variety of conventional fluid samplers (e.g., Ti-majors fluid samplers, gas-tight fluid samplers, and syringe samplers).

The GeoMICROBE sled couples this water-collecting capability with a variety of in-line sensors (e.g., temperature, fluid flow, and in situ electrochemical analyzer; Fig. F11C) (Cowen et al., unpubl. data), including voltammetric measurements with solid-state gold/amalgam (Au/Hg) voltammetric (micro)electrodes (e.g., Luther et al., 2008; Glazer and Rouxel, 2009). In voltammetry, current is measured while scanning a voltage range, allowing simultaneous detection of multiple chemical species (analogous to varying wavelength and measuring absorbance with spectroscopy). Many redox species can be characterized, including O2, Mn2+, Fe2+, H2S/HS, S2O32–, S4O62–, S0, and aqueous FeS and Fe(III) clusters. These in situ measurements are made with an in situ electrochemical analyzer (ISEA) (Analytical Instrument Systems, Inc.). A titanium pressure case houses the analyzer, which consists of a potentiostat, an eight-electrode multiplexer, and an internal computer with 8 GB flash memory. Recent improvements to the ISEA include the capability for simultaneous pH, Eh, temperature, and optical oxygen sensor data collection and a stepper motor to control micromanipulators, valves, and pumps. Waterproof bulkhead connectors (Subconn, Inc.) allow cabled connection to an external power source and RS-232, RS-422, or ethernet communication for real-time measurements (e.g., via submersible) or unattended preprogrammed data collection (e.g., via GeoMICROBE sled). The new ISEA also incorporates a controller that is capable of booting into a stand-alone data logging configuration, utilizing a low-power sleep mode. This dual-boot feature allows the chemical redox environment at a deployment location to be surveyed in real time to determine an optimum voltammetry scanning regime. Additional ongoing developments include miniaturization of analyzer electronics for incorporation into downhole deployment strategies and long-term reliability of electrodes.

Downhole sampling from a submersible

When the initial downhole strings were removed from the boreholes, several boreholes discharged (produced) fluids. This allowed a variety of samplers to be used to collect fluids near the point of discharge (e.g., major and gas-tight samplers). The first sampling effort used a winch and a pump connected to the submersible Alvin to sample the open borehole in Hole 892A. When the data logger and pressure sensor were removed, pieces of hydrate floated out of the borehole and into the water column. Attempts to deploy the sampler failed because the borehole was plugged with gas hydrates, demonstrating the difficulty of conducting a passive, in situ experiment in convergent margin with gas hydrates (Carson et al., 2003).

The Scripps wireline tool coupled with the Gieskes water sampler was used to sample fluids within the borehole in Hole 1026B (Spiess et al., 1992). However, temperature was an issue with the Gieskes water sampler. After exposure (~15 min) to temperatures of 64°C the electronics failed. Instead, fluids were collected from within the borehole using a winch, pump, and sample system deployed by the Alvin, similar to that used in Hole 892A (Wheat et al., 2004). An intake was placed in the borehole, and the winch was used to lower the intake to 175 meters below seafloor (mbsf). A pump mounted on the Alvin brought fluids from depth to the seafloor through a manifold of samplers that were triggered with a manipulator on the submersible. A similar system was deployed in Hole 1200C to collect fluids from the base of the borehole. Here a winch system was deployed with a spring to detect changes in resistance when extracting the sampler, thus minimizing the chance for getting the system caught on a protruding object within the borehole. This microbial sampling device minimized or eliminated contamination with seawater (K. Takai, pers. comm., 2009).